Process indicators (e.g. autoclave tape) must be used for all sterilization processes.
Periodic verification of the sterilization process should be performed every 3 months, as well as after repair.
After any repair, the sterilization process must be verified using biological indicators. The repaired autoclave/dry sterilizer should only be used after the sterilization process has been tested and approved. If using a temporary autoclave/dry sterilizer, the sterilization process must be verified using biological indicators, with satisfactory results before using the equipment.
If using a temporary autoclave/dry sterilizer, the sterilization process must be verified using biological indicators, with satisfactory results before using the equipment.
Please refer to NIR for reprocessing of sterilizable medical equipment, Denmark only: Link, as well as General Practice, Denmark only: Link.
Danske Fodterapeuter have updated and developed their hygiene guidelines, it should help make daily hygiene easier and without doubts.