Recommendations for tattoo artists, podiatrists, and similar industries

Process indicators (e.g. autoclave tape) must be used for all sterilization processes.

It is recommended to control the sterilization process at least once a day using chemical indicator/multivariable chemical indicator.

Periodic verification of the sterilization process should be performed every 3 months, as well as after repair.

After any repair, the sterilization process must be verified using biological indicators. The repaired autoclave/dry sterilizer should only be used after the sterilization process has been tested and approved. If using a temporary autoclave/dry sterilizer, the sterilization process must be verified using biological indicators, with satisfactory results before using the equipment.

If using a temporary autoclave/dry sterilizer, the sterilization process must be verified using biological indicators, with satisfactory results before using the equipment.
Please refer to NIR for reprocessing of sterilizable medical equipment, Denmark only: Link

Safe Sterilization recommends the following subscription plan:

  • C subscription (1 set of biological indicators is sent every 3 months/4 set annually)

When it is time to test, Safe sends you a set of biological indicators. Read more here.